Smart LEDZ Fit
User ManualVer.2.0

  1. Smart LEDZ Fit
    User ManualVer.2.0
  2. Sensor calibration, surface reflectivity, fade time settings

Table of Contents

Sensor calibration, surface reflectivity, fade time settings

Calibrate the sensor and set the surface reflectivity.

In an environment without external light, such as at night, ensure all lighting fixtures are turned on at 100%.

Ensure that the sensor mode and group settings are complete.

*Calibration is not required if you are using the motion sensor and not using the lighting option.

1

Tap A.
The sensor settings screen will open.

2

Make sure there are no people within the sensor’s detection range, then tap B.
A confirmation screen will appear; tap C.

*If you are not using the “Target Illuminance (Lux)” illuminance mode, the calibration is complete at this point.

3

The illuminance directly beneath the sensor is measured by an illuminance sensor, and the surface reflectance is calculated.

Formula:
Surface reflectance = Brightness detection value (D) ÷ Measured illuminance value

4

Tap E.
The sensor pattern settings will open.

5

Tap F.
The surface reflectivity settings screen will open.

6

Enter the surface reflectance calculated in step 3 into G either as a numerical value or by adjusting the bar.

7

Tap H.
Surface reflectance setting is complete.

Sensor fade time setting

1

Tap A.
The sensor settings screen will open.

2

Tap I.
The fade settings screen will open.
Scroll J to set the minutes and seconds as desired.
Tap K.
The fade settings are complete.